CONFIGURING
THE METER
USING
SETPOINT 1 CONFIGURATION ("S1.CF")
Setpoint 1 Configuration (S1.CF) is not active unless your meter
has dual relay output capabilities. The LED's will display whether the
("S1.CF") is active or not. You may use Setpoint 1 Configuration
("S1.CF") for the following:
* Setting setpoint's active band above or below your chosen value
* Selecting whether the setpoint operation is latched or unlatched
* Assigning setpoint values to the net or gross reading
Setting Setpoint 1's Active Band
1. Press the MENU button until the meter shows "S1.CF".
2. Press the /TARE
button. The meter shows one of the following:
* "S.1=A" (Active above the deadband) (default)
* "S.1=B" (Active below the deadband)
3. Press the /NT/GRS
button to toggle between available choices.
4. Press the /TARE
button to select if Setpoint 1 is latched or unlatched or press the MENU
button to store your selection and enter "S2.CF" (Setpoint 2 Configuration)
Selecting if Setpoint 1 is Latched or Unlatched
1. Press the /TARE
button. The meter shows one of the following:
* "S.2=L" (setpoint 1 to be latched) (default)
* "S.2=U" (setpoint 1 to be unlatched)
2. Press the /NT/GRS
button to toggle between available choices.
3.
Press the /TARE
button to assign Setpoint 1 values to net or gross reading or press the
MENU button to enter "S2.CF" (Setpoint 2 Configuration).
Assigning Setpoint 1 Values to Net or Gross Readings
1. Press the /TARE
button. The meter shows one of the following:
* "S.3=N" setpoint 1 assigned to net reading (default)
* "S.3=G" setpoint 1 assigned to gross reading
2. Press the /NT/GRS
button to toggle between available choices.
3. Press the MENU button to store your selection(s). The
meter momentarily shows "STRD", followed by "S2.CF" (Setpoint
2 Configuration).
USING SETPOINT 2 CONFIGURATION
("S2.CF")
Setpoint 2 Configuration ("S2.CF") is not active unless your meter
has dual relay output capabilities. The LED's will display whether the
("S2.CF") is active or not. You may use Setpoint 2 Configuration
("S2.CF") for the following:
* To set setpoint's
active band above or below your chosen value
* To select whether the setpoint operation is latched or unlatched
* To assign setpoint values to the net or gross reading
Setting
Setpoint 2's Active Band
1. Press the MENU button until the meter shows "S2.CF".
2.
Press the /TARE
button. The meter shows one of the following:
* "S.1=A" (Active above the deadband) (default)
* "S.1=B" (Active below the deadband)
3. Press the /NT/GRS
button to toggle between available choices.
4. Press the /TARE
button to select if Setpoint 1 is latched or unlatched or press the MENU
button to store your selection and enter "S1.DB" (Setpoint 1 Deadband)
Selecting if Setpoint 2 is
Latched or Unlatched
1. Press the /TARE
button. The meter shows one of the following:
* "S.2=L" (setpoint 1 to be latched) (default)
* "S.2=U" (setpoint 1 to be unlatched)
2. Press the /NT/GRS
button to toggle between available choices.
3. Press the /TARE
button to assign Setpoint 2 values to net or gross reading or press the
MENU button to enter "S1.DB" (Setpoint 1 Deadband).
Assigning Setpoint 2 Values
to Net or Gross Readings
1. Press the /TARE
button. The meter shows one of the following:
* "S.3=N" (setpoint 2 assigned to net reading)
(default)
* "S.3=G" (setpoint 2 assigned to gross reading)
2. Press the /NT/GRS
button to toggle between available choices.
3. Press the MENU button to store your selection(s). The
meter momentarily shows "STRD", followed by "S1.DB" (Setpoint
1 Deadband).
SETTING THE SETPOINT 1 DEADBAND
("S1.DB")
Setpoint 1 Deadband ("S1.DB") is not active unless your meter
has dual relay output capabilities. The LED's will display whether the
("S1.DB") is active or not. The Setpoint 1 default deadband is
0003. To change the deadband (hysteresis) of setpoint 1, follow these
steps:
1. Press the MENU button until the meter shows "S1.DB".
2. Press the /TARE
button. The meter shows the last previously stored 4-digit number (0000
through 9999) with flashing 4th digit.
3. Press the /NT/GRS
button to change the value of the flashing digit. If you continue to press
the /NT/GRS
button, the flashing digit's value continues to change.
4. Press the /TARE
button to scroll horizontally to the next digit.
5. Press the MENU button to store your selection. The meter
momentarily shows "STRD", followed by "S2.DB" (Setpoint
2 Deadband).
SETTING
THE SETPOINT 2 DEADBAND ("S2.DB")
Setpoint 2 Deadband ("S2.DB") is not active unless your meter has
dual relay output capabilities. The LED's will display whether the (S2.DB)
is active or not. The Setpoint 2 default deadband is 0003. To change the
deadband (hysteresis) of setpoint 2, follow these steps:
1. Press the MENU button until the meter shows "S2.DB".
2. Press the /TARE
button. The meter shows the last previously stored 4-digit number (0000
through 9999) with flashing 4th digit.
3. Press the /NT/GRS
button to change the value of the flashing digit. If you continue to press
the /NT/GRS
button, the flashing digit's value continues to change.
4. Press the /TARE
button to scroll horizontally to the next digit.
5. Press the MENU button to store your selection. The meter
momentarily shows "STRD", followed by "OT.CF" (Output Configuration).
|